ACCURACY OF COMPRESSED SENSING ACCELERATED 3D-FLAIR IN THE DIAGNOSTIC WORK-UP OF MULTIPLE SCLEROSIS AT 1.5 T
medRxiv · · Preprint · Open access
Iacoban, Klüner + 6 more
Abstract ↗AI summary
The abstract is read at the publisher; the summary is JClub's.
Compressed sensing–accelerated 3D-FLAIR at 1.5T detects multiple sclerosis lesions non-inferiorly to optimized GRAPPA while cutting scan time by 31.8% (4:00 vs 5:52 minutes).
- Why it matters: Long 3D-FLAIR acquisition times limit MR-based MS diagnosis and monitoring, creating a need to shorten scans without losing lesion detection performance.
- What they did: In a single-center prospective non-inferiority study, optimized CS- and GRAPPA-accelerated 3D-FLAIR were compared in 68 patients with known or suspected MS, using manual lesion counts by two blinded readers with a 10% margin.
- The result: Non-inferiority for manual lesion count was shown for both readers (p < .0001 and p = .005), with excellent interrater agreement for CS (ICC = 0.921) and very good for GRAPPA (ICC = 0.851).
The findingWhy it mattersWhat they didThe result
AI summary of the abstract · may contain errors · not medical advice · read the paper.