For a 1045 nm excitation laser in a dye cuvette, fluorescence maxima at the sample occurred between 21 and 24 screw turns in 11 of 13 sweeps.
- Why it matters: Two-photon absorption efficiency depends on compensating dispersion across the entire excitation path, and existing fluorescence-based adjustment needs a practical, sample-referenced workflow.
- What they did: The method uses side-view two-photon fluorescence imaging of a dye cuvette, with reference records, stability warnings, and open analysis software, independent of microscope scanning and detection.
- The result: For a 920 nm laser, sample fluorescence maxima differed by 1.63 kfs^2 on average from autocorrelation minima before the scan optics, while recordings reveal limits from instability and clipping.