EEG decoding during transcranial focused ultrasound (tFUS) to human primary visual cortex (V1) is dominated by a mechanical, pulse-locked artifact, reaching up to 93% accuracy for pulse duration.
- Why it matters: It is unclear whether commonly reported EEG peaks during tFUS reflect neuronal engagement or non-neuronal signals produced by the ultrasound and its vibration.
- What they did: In eight participants receiving 150- or 300-ms pulses or sham, EEG decoding distinguished tFUS from sham up to 93% for exactly each pulse duration, but the same pulse-locked step appeared across all channels and scaled with drive, also appearing in a foam head and a forehead piezo pickup.
- The result: Because the sonication-locked decoding signal was identified as mechanical, EEG peaks within a pulse cannot by themselves be taken as evidence of cortical engagement, motivating active shams that reproduce vibration without focusing.