X-ray hydroxyl radical footprinting monitored by mass spectrometry (XF-MS) quantitatively reproduces known protein thermodynamic stability values using ~500-fold less protein than circular dichroism.
- Why it matters: Traditional stability assays need large amounts of pure protein in ideal conditions, limiting measurements in heterogeneous or limited samples and preventing local, domain-level insight.
- What they did: The authors combined XF-MS with chemical denaturation, first validating agreement with established stability values, then applying peptide-level XF-MS to heterogeneous mixtures including cell lysate.
- The result: XF-MS enables global and local stability measurements, detecting deviations from two-state unfolding and assigning thermodynamic values to individual structural domains in multi-domain proteins.
The findingWhy it mattersWhat they didThe result