In 26 hospitalized daily tobacco smokers monitored for over 40 hours, Vitalograph BreathCO reached <3 ppm in 100% of participants, defining a key abstinence benchmark.
- Why it matters: Breath CO is widely used to verify smoking outcomes, but the natural multi-day decline under forced abstinence and device-to-device cutoff differences were not well characterized.
- What they did: Over 44 hours on an inpatient unit, four monitors (Vitalograph BreathCO, piCO+, iCOquit corded and Bluetooth) were compared using linear mixed models and ROC analyses with logistic regression.
- The result: Using Vitalograph BreathCO as the reference, optimal cutoffs were <3 ppm for both iCOquit versions and <5 ppm for piCO+, and devices showed non-interchangeable thresholds.