Raw photoplethysmography from low-cost wearable sensors shows a reproducible exponential decay after electrically induced ventricular fibrillation in 9 pigs, with a 10.53 s time constant.
- Why it matters: Unwitnessed out-of-hospital cardiac arrest makes time to resuscitation critical, and existing wearable systems may trade specificity for sensitivity.
- What they did: Invasively monitored pigs underwent induced ventricular fibrillation while raw PPG was recorded; a heuristic algorithm was tested with leave-one-out cross-validation for sensitivity and latency, and assessed for false positives using 57,857 hours of free-living PPG from 115 participants.
- The result: The algorithm detected ventricular fibrillation in 8 of 9 pigs (sensitivity 0.889; 95% CI, 0.518-0.997) with a median latency of 57.2 s, and generated 2.73 false positives per user-year (95% CI, 1.62-4.31) in humans.