A collagen hybridizing peptide workflow in porcine aorta detects load-induced collagen damage before failure, with damage scores comparable at 1.2× stretch and elevated beyond 1.2×.
- Why it matters: Quantifying pre-failure injury during endovascular device deployment has been largely unfeasible, limiting validation and optimization of endovascular aortic devices.
- What they did: The team optimized an ex vivo large-artery protocol for energy-storage and geometry, selecting a CHP fluorophore by measuring tissue autofluorescence and using tensile testing plus fluorescence/confocal imaging to score damage versus stretch.
- The result: Collagen damage increased only beyond 1.2× stretch, matching higher complication rates with excessive stent oversizing, and enabling adaptation to ex vivo stent deployment comparisons and constitutive modeling.