Temporally aware reconstruction localized robotically manipulated endovascular devices in an anthropomorphic phantom from retrospectively sparsified pulsed sparse-view CT, with median tip-to-vessel-wall error 0.03 mm at 6 projections.
- Why it matters: Endovascular fluoroscopy lacks depth information, and sparse projection CT could enable volumetric updates without full acquisition, but accuracy for device localization must be shown.
- What they did: A modular phantom was scanned with 1248 projections, then retrospectively sparsified to 6, 12, 24, 32, and 52 projections; filtered backprojection and a temporally aware prior-image method were compared against full-view filtered backprojection.
- The result: True pulsed and matched retrospectively sparsified images had CT number differences of 4.8–6.2 Hounsfield units with corrected standard deviations of 27.5–29.9, and temporally aware reconstruction depicted both devices at all projection counts.