In 30 healthy participants during oral glucose tolerance tests, mid-infrared optoacoustic sensing produced an Optoacoustic Mid-infrared Signal with MAE 11.9 mg/dL versus reference blood glucose.
- Why it matters: CGM currently uses needle-based interstitial measurements that lag behind blood glucose, and MiROS has only shown glucose sensitivity in mice, leaving human performance unknown.
- What they did: The pilot study defined OMS as a readout of glucose-linked tissue optical-property changes and compared paired OMS values with blood glucose during oral glucose tolerance tests.
- The result: OMS achieved MAPE 10% and placed more than 99.9% of paired values in consensus error grid zones A and B, and it compared favorably with CGM (MAE 15.7 mg/dL, MAPE 12%) without a comparable time lag.