Deep learning models estimating invasive intracranial pressure waveforms from extracranial ABP, PPG, and ECG achieved external MAE of 4.266–5.899 mmHg and r of 0.585–0.719 in 158 critically ill adults.
- Why it matters: Invasive ICP monitoring is limited to a subset of critically ill patients, so a noninvasive waveform estimate could broaden access while preserving clinically relevant accuracy.
- What they did: Five deep learning architectures were trained on ABP, PPG, and ECG waveforms using invasive intraparenchymal ICP as ground truth, then externally validated on the MIMIC-III Waveform Database with MAE and Pearson correlation.
- The result: Three of five architectures met a clinically motivated 5 mmHg MAE benchmark, generalized across institutions for mean pressure and pulsatile trends, but did not reproduce beat-level peak morphology.
The findingWhy it mattersWhat they didThe result