Low-field (0.55T) bi-parametric prostate MRI with PI-RADS–compliant parameters achieved clinically acceptable diagnostic quality in 21/22 patients, with patient-level lesion sensitivity of 91% versus 3T.
- Why it matters: Low-field prostate MRI could expand access, but feasibility and lesion detection performance relative to standard 3T needed validation using PI-RADS–aligned acquisition.
- What they did: In 15 healthy men and 22 suspected prostate cancer patients, tri-planar T2-weighted imaging and axial diffusion-weighted imaging were acquired at 0.55T with deep learning reconstruction; DWI used b=50 and 800 s mm-2 (calculated 1500 s mm-2), and 3T served as the reference for lesion detection.
- The result: DLR improved T2-WI quality in controls (p<0.05), and low-field patient-level sensitivity was 91% for both readers without significant difference from 3T (p=0.5), supporting clinical feasibility.