bioRxivJClub
Microfluidic sieve and detector for rapid ultrasensitive assays with single-molecule sensitivity
bioRxiv · · Preprint · Open access
Kim, Shafique + 3 more
Abstract ↗AI summary
The abstract is read at the publisher; the summary is JClub's.
The microfluidic sieve and detector (MFSD) captures and quantifies analytes from 200 microlitres in one minute and four minutes total, reaching zeptomolar (10^-19M) sensitivity.
- Why it matters: Ultrasensitive assays still take hours to deliver results, limiting clinical and research impact. MFSD targets faster, single-molecule–level readout with predictable capture under flow.
- What they did: MFSD uses half a million 8-m-diameter pores that bind analytes via Brownian motion-induced wall collisions, guided by ensemble and stochastic Brownian agent dynamics. It completes digital sandwich assays by sequentially flowing sample and reagents, partitioning pores, and enzymatically revealing single trapped analytes.
- The result: MFSD tests are specific and reproducible with a large dynamic range, measuring proteins including influenza A nucleoprotein in clinical samples and nucleic acids. It is faster and more sensitive than PCR, enabling rapid ultrasensitive diagnostics.
The findingWhy it mattersWhat they didThe result
AI summary of the abstract · may contain errors · not medical advice · read the paper.
- Open access