Distortion-minimization validation for optically pumped MEG is tolerant to unmodelled sensor errors—under 2 mm per axis, 5° orientation, 10° array displacement, and 10% gain—when simulation and inversion match.
- Why it matters: Because cortical current-flow inversion is ill-posed, head-cast–style validation needs constraints for OP-MEG when key errors (sensor geometry and gain) are not modeled.
- What they did: Simulations of cortical current flow were contaminated with sensor position/orientation errors and gain changes, then distortion was minimized only when simulation and inversion assumptions matched; empirical error magnitudes were quantified in mm relative to the true cortical mesh.
- The result: With no unmodelled errors, distortion minimized under matched assumptions, and empirical OPM datasets showed that reconstructions using minimum norm and beamformer assumptions were sensitive to underlying anatomy.