SEA-Lab achieved 90.41% protocol-review pass rates across 219 tasks and raised virtual-validation pass rates on 66 automation-eligible tasks from 16.7% to 87.7% after two revisions.
- Why it matters: Reliable lab execution depends on turning AI-generated protocols into procedures that fit local equipment, while preserving execution feedback so future planning can improve.
- What they did: The system used a Dry-Lab Harness to generate machine-readable protocols, a wet-lab execution harness to coordinate heterogeneous devices and log traces, and recursive self-improvement to revise execution policies using physical outcomes.
- The result: On two liquid handlers, runtime dropped by 45.8% for plate replication, 34.1% for well rearrangement, and up to 68.6% for sample normalization; in a phage-display workflow, 93 of 96 samples produced sequence-correct clones and ELISA screening found ten candidates.