AI-guided handheld echocardiography by nurse practitioners in rural Damak, Nepal detected any valvular lesion with 95.8% sensitivity, 100% specificity, and kappa 0.97 versus cardiologist overread.
- Why it matters: Rural Nepal has limited access to echocardiography and cardiovascular specialists, leaving rheumatic heart disease screening constrained and potentially delaying early detection.
- What they did: In November 2025, nurse practitioners used EchoNous Kosmos Plus devices for community screening camps, with cardiologist overread as the comparator; diagnostic accuracy used sensitivity, specificity, and kappa, alongside semi-structured interviews.
- The result: Scan time averaged 5.8 minutes, and lesion-specific accuracy ranged from 66.7% sensitivity for mitral regurgitation to 100% sensitivity for mitral stenosis and tricuspid regurgitation; feasibility was supported by portability and remote review but limited by training, workflow constraints, and technical issues.
The findingWhy it mattersWhat they didThe result